FireLite Furnace Operations Manual Version 1.00.7 – 1/05
Fast Cool with Vacuum
The FireLite comes from the factory with a Fast Cool
Vacuum feature. This feature automatically turns on the
vacuum pump between programs to cool the muffle to the
entry temperature of the new program.
To use this feature:
1. Press the Start Program button.
Note: If the final temperature of the last program is higher
than the start temperature of the new program, the vacuum
pump will automatically come on.
To deactivate this feature:
1. Press the Special Function button.
2. Press 82 on the keypad.
3. Press the Enter button.
4. Press 0 (zero) to change the display from YES to NO
5. Press the Enter button.
To reactivate this feature:
1. Press the Special Function button.
2. Press 82 on the keypad.
3. Press the Enter button.
4. Press any number key between 1 and 9 to change the
four-digit display from NO to YES.
5. Press the Enter button.
Adjusting Calibrations for Low-Fusing
and High-Fusing Porcelains
All porcelain furnaces fire slightly differently. Even porcelain
furnaces from the same manufacturer can fire porcelains at
slightly different temperatures. The FireLite addresses this
problem by allowing you to customize the calibration. This
feature can be used to adjust your furnace to fire porcelains
at manufacturers’ recommendations, or to adjust multiple
furnaces to fire identically.
The FireLite calibration can be adjusted for normal high-
fusing porcelains and for the new low-fusing porcelains.
These adjustments do not affect each other, so both
adjustments can be used if you fire both types of porcelain.
The adjustments can range from 0° C/32°F to 50° C/90° F.
Adjust High-Fusing Porcelain Calibration
(800° C/1,470° F or above)
Overfires:
1. Press the Special Function button.
2. Press 18 on the keypad.
3. Press the Enter button.
4. Enter the number of degrees to adjust the
calibration downward.
5. Press the Enter button.
Underfires:
1. Press the Special Function button.
2. Press 19 on the keypad.
3. Press the Enter button.
4. Enter the number of degrees to adjust the
calibration upward.
5. Press the Enter button.
Adjust Low-Fusing Porcelain Calibration
(800° C/1,470° F or below)
Overfires:
1. Press the Special Function button.
2. Press 16 on the keypad.
3. Press the Enter button.
4. Enter the number of degrees to adjust the
calibration downward.
5. Press the Enter button.
Underfires:
1. Press the Special Function button.
2. Press 17 on the keypad.
3. Press the Enter button.
4. Enter the number of degrees to adjust the
calibration upward.
5. Press the Enter button.